Beckman Coulter Inc. recalls Lipase, Catalog No. OSR6x30 Product Usage: Lipase OSR6x30 is intended as a System reagent for the quantitative determin…
- Recall date
- April 13, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1930-2016
- FDA classification
- Class II
- Brand / firm
- Beckman Coulter Inc.
- Sold / distributed
- Worldwide Distribution - US Nationwide including Puerto Rico and the countries of: Albania, Algeria, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denma…
Why it was recalled
Beckman Coulter has identified that Dipyrone (Metamizole) may cause erroneous, false low results in the following assays: Cholesterol OSR6X16, Triglyceride OSR6X118, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Lipase, Catalog No. OSR6x30 Product Usage: Lipase OSR6x30 is intended as a System reagent for the quantitative determination of Lipase activity in human serum on Beckman Coulter AU analyzers.