Beckman Coulter, Inc. recalls MicroScan Pos Combo Panel Type 33 (PC33): Catalog number B1017-211 MicroScan Pos Combo 33 B1017-211 panels are designed…
- Recall date
- April 8, 2014
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1524-2015
- FDA classification
- Class II
- Brand / firm
- Beckman Coulter, Inc.
- Sold / distributed
- Nationwide Distribution including Puerto Rico and to the states of : CT, GA, NC, OK, NY, SC, KY, TN, CA, WA, OR, NJ, OH, PA, FL, MI, TX, WI, MO and NH.
Why it was recalled
Siemens investigation confirmed that a portion one lot of panel Pos Combo 33 lot was packaged without desiccants. The substrates on a panel packaged without a desiccant may deteriorate over time due to exposure to additional moisture in the panel pouch.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
MicroScan Pos Combo Panel Type 33 (PC33): Catalog number B1017-211 MicroScan Pos Combo 33 B1017-211 panels are designed for use in determining quantitative and /or qualitative antimicrobial agent susceptibility and/or identification to the species level of colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic gram-positive bacteria.