Medical device recalls Moderate risk

Beckman Coulter Inc. recalls Salicylate Reagent (SALY 2 x 45), Catalog No. 378194. Intended for quantitative determination of salicylate concentrati…

Recall date
April 3, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2084-2017
FDA classification
Class II
Brand / firm
Beckman Coulter Inc.
Sold / distributed
Worldwide Distribution -- Australia, Bangladesh, Belgium, Brunei, Bulgaria, Canada, China, Colombia, Czech Republic, Egypt, Eritrea, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Net…

Why it was recalled

Beckman Coulter has identified the following drugs as potential sources of interference: Metamizole (Dipyrone), Sulfasalazine, and Sulfapyridine and may result in false low results.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Salicylate Reagent (SALY 2 x 45), Catalog No. 378194. Intended for quantitative determination of salicylate concentration in human serum or plasma.