Medical device recalls Moderate risk

Beckman Coulter Inc. recalls Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K , Ref. 506-30. The firm names on the label are Sekisui Diagnosti…

Recall date
July 20, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-3118-2017
FDA classification
Class II
Brand / firm
Beckman Coulter Inc.
Sold / distributed
Distribution US nationwide.

Why it was recalled

Reagent carryover from Sekisui ACTH when utilized on the BEC AU 680 analyzer may lead to falsely elevated BEC Lipase recovery exceeding the total precision specification.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K , Ref. 506-30. The firm names on the label are Sekisui Diagnostics (UK) Limited, United Kingdom, and Sekisui Diagnostics P.E.I. Inc., Canada. Used in laboratory testing.