Medical device recalls Moderate risk

Beckman Coulter Inc. recalls UniCel DxH 800 Coulter Cellular Analysis System, Catalog No. 629029, B24465, B24802, B66445, B68304, B63322.

Recall date
April 24, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2157-2017
FDA classification
Class II
Brand / firm
Beckman Coulter Inc.
Sold / distributed
Worldwide distribution: US nationwide, Algeria, Andorra, Austria, Australia, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Canada, China, Colombia, Costa Rica, Cote d'lvoire, Croatia, Czech Republic, Ecuador, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, I…

Why it was recalled

There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual light scatter pattern may be incorrectly removed from analysis which can result in an erroneous differential result.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

UniCel DxH 800 Coulter Cellular Analysis System, Catalog No. 629029, B24465, B24802, B66445, B68304, B63322.