Medical device recalls Moderate risk

Becton, Dickinson and Company, BD Biosciences recalls BD Trucount Tubes (Cat. No. 340334), used for determining absolute counts of leucocytes in blood. BD Trucount tubes are…

Recall date
October 3, 2022
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0171-2023
FDA classification
Class II
Brand / firm
Becton, Dickinson and Company, BD Biosciences
Sold / distributed
U.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, VT, WA, and WI O.U.S.: Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Dominican Republic…

Why it was recalled

Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant. Firm root cause investigation confirmed that label detachment is due to an adhesive change by a tier 2 supplier.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

BD Trucount Tubes (Cat. No. 340334), used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used with the BD FACS Loader.

Get recall alerts

Free email alert whenever Becton, Dickinson and Company, BD Biosciences has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Becton, Dickinson and Company, BD Biosciences