Medical device recalls Moderate risk

Becton Dickinson & Company recalls BD NEXIVA Closed IV Catheter System Dual Port 18GA 1.25 IN (BD Nexiva Catheter), Catalog Number 383539 Product Usage: B…

Recall date
August 23, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0384-2019
FDA classification
Class II
Brand / firm
Becton Dickinson & Company
Sold / distributed
Worldwide Distribution -US Nationwide in the states of: CA, CO, FL, GA, IL, LA, MI, ND, NE, NY, OH, PA, RI, SC, VA, WA and WV. The products were distributed to the following foreign countries: Argentina, Australia, Belgium, Brazil, Canada, Colombia, Costa Rica, Mexico and New Zealand.

Why it was recalled

One lot of the specified needles is not performing as intended due to a failure in the needle tip shield/safety mechanism, which may lead to an increased risk of a contaminated needlestick injury.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

BD NEXIVA Closed IV Catheter System Dual Port 18GA 1.25 IN (BD Nexiva Catheter), Catalog Number 383539 Product Usage: BD Nexiva Closed IV Catheter System Dual Port (hereafter referred to as BD Nexiva device or catheter) (see Figure below) incorporates a passive needle shielding mechanism that is activated as part of the usual insertion procedure when the needle is withdrawn from the catheter. BD Nexiva device is designed to protect the clinician from accidental needle stick injuries and minimize exposure to blood during the insertion process. The needle shield is designed to not release from the catheter adapter until the needle tip is protected and cannot be rearmed once activated. The BD Nexiva device is intended for use in the peripheral vascular system to provide access for the administration of fluids, medication, blood and blood products, and for the withdrawal of blood samples. This catheter may be used for any patient population with consideration given to patient size, appropriateness for solutions being infused and duration of therapy.

Get recall alerts

Free email alert whenever Becton Dickinson & Company has a new recall — straight from official government data. Unsubscribe anytime.