Becton Dickinson & Company recalls BD Vacutainer¿ EDTA Tan Top Tube; Catalog Number: 367855 (UDI: 50382903678551)
- Recall date
- March 22, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1140-2018
- FDA classification
- Class I
- Brand / firm
- Becton Dickinson & Company
- Sold / distributed
- Distributed in 46 states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Puerto Rico.
Why it was recalled
BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
BD Vacutainer¿ EDTA Tan Top Tube; Catalog Number: 367855 (UDI: 50382903678551)