Becton Dickinson & Company recalls BD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 2.0 mL (Pink) BD Hemogard Additive: K2EDTA (spray dried), 3.6mg…
- Recall date
- December 3, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0567-2016
- FDA classification
- Class II
- Brand / firm
- Becton Dickinson & Company
- Sold / distributed
- Worldwide Distribution: US (Nationwide) and the countries of: European Union, Peoples Republic of China, Japan and Republic of Korea.
Why it was recalled
Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
BD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 2.0 mL (Pink) BD Hemogard Additive: K2EDTA (spray dried), 3.6mg Sterile IVD BD, Made in USA The Vacutainer Brand PLUS (Plastic) Tube with EDTA Vacutainer¿ Brand Serum Tube are evacuated blood collection tubes which provide a means of collecting, transporting, separating and processing blood in a plastic tube.
Get recall alerts
Free email alert whenever Becton Dickinson & Company has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Becton Dickinson & Company