Bio-Rad Laboratories Inc recalls PR4100 Microplate Reader US, Catalog Number 45591587 Product Usage: The PR4100 Microplate Reader is a 96 well absorbanc…
- Recall date
- April 2, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2541-2018
- FDA classification
- Class II
- Brand / firm
- Bio-Rad Laboratories Inc
- Sold / distributed
- US Nationwide Distribution inn the states of : AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OR, PA, SC, TN, TX, VA, VT, WA, WI, and Puerto Rico.
Why it was recalled
The connector of the power supply was capable of fitting other components provided with the system. However, the voltage output of the reader power supply is incompatible with the other component and, if incorrectly connected, could result in irreparable damage to the system.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
PR4100 Microplate Reader US, Catalog Number 45591587 Product Usage: The PR4100 Microplate Reader is a 96 well absorbance reader for the measurement of light absorbance (optical density-OD) of liquid media. The instrument is intended to be used primarily for the in vitro diagnostic analysis of samples from the human body to obtain information oh physiological and pathological states.
Get recall alerts
Free email alert whenever Bio-Rad Laboratories Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Bio-Rad Laboratories Inc