Biocare Medical, LLC recalls intelliPATH Universal HRP Detection Kit, REF: IPK5011G80, containing DAB Buffer, Catalog Number: IPBF5009G20; ONCORE Pr…
- Recall date
- September 25, 2024
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0315-2025
- FDA classification
- Class II
- Brand / firm
- Biocare Medical, LLC
- Sold / distributed
- US Nationwide distribution in the states of MA, AZ, CT, NY, FL, NC, IN, IL, MO, CO, TX, CA, MI, PR, PA, MD, AL, OH, KY.
Why it was recalled
Kits containing the DAB chromogen buffer, intended for use in either manual or automated Immunohistochemistry (IHC) staining protocols, may produce weak to no staining, which may lead to test failure requiring the test to be repeated.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
intelliPATH Universal HRP Detection Kit, REF: IPK5011G80, containing DAB Buffer, Catalog Number: IPBF5009G20; ONCORE Pro DAB Chromogen, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60; IP DAB Chromogen Kit 80 ml, REF: IPK5010G80, containing DAB Buffer, Catalog Number: IPBF5009G20; DAB kit, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60 Folate Receptor alpha IHC Assay Kit, REF: IPI4006KG10, containing DAB Buffer, Catalog Number: IPBF5009G20
Get recall alerts
Free email alert whenever Biocare Medical, LLC has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Biocare Medical, LLC