Medical device recalls Moderate risk

FILMARRAY TORCH MODULE BOX recalled over Salmonella risk

Recall date
August 7, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Official notice title
BioFire Diagnostics, LLC recalls FILMARRAY TORCH MODULE BOX, HTFA-ASY-0003, HTFA-ASY-0003-W, HTFA-ASY-0003U
Recall number
Z-3123-2024
FDA classification
Class II
Brand / firm
BioFire Diagnostics, LLC
Sold / distributed
Worldwide - US Nationwide distribution including in the states of NC, TX, OH, CA, KS, AL, WA, NY, PA, IL, WI, KY, LA, NE, FL, UT, IA, SC, VA, AZ, CO, WV, MO, TN, MN, DC, DE, OK, WY, IN, NJ, NM, ID, OR, MS, SD, MI, MT, GA, HI, NV, MD, MA, AR, NH, PR, ND, VT, CT, AK, ME, RI and the countries of Korea…

Why it was recalled

Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting temperatures, which may lead to erroneous results including false positive Salmonella, Giardia lamblia, false positive/negative Norovirus, and false negative MRSA.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

FILMARRAY TORCH MODULE BOX, HTFA-ASY-0003, HTFA-ASY-0003-W, HTFA-ASY-0003U