Medical device recalls High risk

BioMedical Equipment Service Co (BMES) recalls Alaris Infusion Pump Module, Model 8100, serviced/repaired by Biomedical Equipment Service Co. (BMES) with an impacted…

Recall date
February 25, 2021
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2204-2021
FDA classification
Class I
Brand / firm
BioMedical Equipment Service Co (BMES)
Sold / distributed
The affected products were distributed to the following US states: CA, IL, IN, KY, LA, NJ, NV, OH, and PA.

Why it was recalled

Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

Alaris Infusion Pump Module, Model 8100, serviced/repaired by Biomedical Equipment Service Co. (BMES) with an impacted bezel repair part TIPA-8100-4410

Get recall alerts

Free email alert whenever BioMedical Equipment Service Co (BMES) has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: BioMedical Equipment Service Co (BMES)