BioMedical Equipment Service Co (BMES) recalls Alaris Infusion Pump Module, Model 8100, serviced/repaired by Biomedical Equipment Service Co. (BMES) with an impacted…
- Recall date
- February 25, 2021
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2204-2021
- FDA classification
- Class I
- Brand / firm
- BioMedical Equipment Service Co (BMES)
- Sold / distributed
- The affected products were distributed to the following US states: CA, IL, IN, KY, LA, NJ, NV, OH, and PA.
Why it was recalled
Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
Alaris Infusion Pump Module, Model 8100, serviced/repaired by Biomedical Equipment Service Co. (BMES) with an impacted bezel repair part TIPA-8100-4410
Get recall alerts
Free email alert whenever BioMedical Equipment Service Co (BMES) has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: BioMedical Equipment Service Co (BMES)