VITEK¿ 2 Compact 15 recalled over mold contamination
- Recall date
- July 12, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Biomerieux Inc recalls VITEK¿ 2 Compact 15, REF numbers: 27415 and 27415R software and reagent cards designed for the identification (ID) and…
- Recall number
- Z-3075-2017
- FDA classification
- Class II
- Brand / firm
- Biomerieux Inc
- Sold / distributed
- Worldwide
Why it was recalled
Customers have reported that some VITEK¿ 2 cards are staying in preliminary status, not finalizing after ejection from the instrument, and not allowing cards in subsequent carousel slots to be processed. The issue was reported to occur on VITEK¿ 2 Compact 15 and Compact 30 systems following a system software update to version 8.01.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
VITEK¿ 2 Compact 15, REF numbers: 27415 and 27415R software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing (AST) of bacteria and yeast.