Medical device recalls Moderate risk

VITEK¿ 2 Compact 15 recalled over mold contamination

Recall date
July 12, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Official notice title
Biomerieux Inc recalls VITEK¿ 2 Compact 15, REF numbers: 27415 and 27415R software and reagent cards designed for the identification (ID) and…
Recall number
Z-3075-2017
FDA classification
Class II
Brand / firm
Biomerieux Inc
Sold / distributed
Worldwide

Why it was recalled

Customers have reported that some VITEK¿ 2 cards are staying in preliminary status, not finalizing after ejection from the instrument, and not allowing cards in subsequent carousel slots to be processed. The issue was reported to occur on VITEK¿ 2 Compact 15 and Compact 30 systems following a system software update to version 8.01.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

VITEK¿ 2 Compact 15, REF numbers: 27415 and 27415R software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing (AST) of bacteria and yeast.