VITEK 2 GP recalled over mold contamination
- Recall date
- August 25, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Biomerieux Inc recalls VITEK(R) 2 GP, REF 21342, IVD, UDI 03573026131920, bioMerieux, Inc. VITEK(R) 2 is an automated system consisting of ins…
- Recall number
- Z-3134-2017
- FDA classification
- Class II
- Brand / firm
- Biomerieux Inc
- Sold / distributed
- US and Angola, Argentina, Austria, Belarus, Belgium, Burkina Faso, Cameroon, Canada, China, Congo, Croatia, Egypt, France, Germany, Greece, Guam, Hungary, Israel, Italy, Latvia, Luxembourg, Macedonia, Mali, Mexico, Morocco, Netherlands, Nicaragua, Oman, Panama, Poland, Saudi Arabia, Serbia, South A…
Why it was recalled
Customer reports have indicated occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC(R) 700327(TM) in association with VITEK(R) 2 GP ID Lot 2420240403 and 2420303403, causing failure of the Quality Control (QC) testing for the VITEK(R) 2 GP ID card. The expected O129R reaction for ATCC¿ 700327(TM) strain is positive.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
VITEK(R) 2 GP, REF 21342, IVD, UDI 03573026131920, bioMerieux, Inc. VITEK(R) 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK¿ 2 utilizes growth based biochemical patterns to determine identification. The VITEK¿ 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria