Medical device recalls Moderate risk

BioMerieux SA recalls VIDAS TOXO IgG Avidity is an automated qualitative test for use on the VIDAS family instruments, for the determination…

Recall date
March 23, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2169-2018
FDA classification
Class II
Brand / firm
BioMerieux SA
Sold / distributed
International distribution.

Why it was recalled

A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

VIDAS TOXO IgG Avidity is an automated qualitative test for use on the VIDAS family instruments, for the determination of anti-toxoplasma IgG avidity in human serum or plasma (lithium heparin, sodium citrate, EDTA) using the ELFA technique (Enzyme Linked Fluorescent Assay). Toxoplasma gondii serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Toxoplasma gondii in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Toxoplasma gondii from clinical specimens.

Get recall alerts

Free email alert whenever BioMerieux SA has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: BioMerieux SA