Medical device recalls Moderate risk

BIOMET FRANCE S.A.R.L. recalls The Optigun Ratchet is a hand-held cement gun for use with Optivac. The Optigun Ratchet is used to deliver bone cement…

Recall date
May 14, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1744-2015
FDA classification
Class II
Brand / firm
BIOMET FRANCE S.A.R.L.
Sold / distributed
Nationwide distribution to MD only

Why it was recalled

The pin which maintains the knob button, may disconnect and become lost. This may result in a delay of surgery to obtain a new Optigun and detect/retrieve the pin, and may necessitate revision surgery if the pin falls into the surgical site and is not detected.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

The Optigun Ratchet is a hand-held cement gun for use with Optivac. The Optigun Ratchet is used to deliver bone cement from the Optivac.

Get recall alerts

Free email alert whenever BIOMET FRANCE S.A.R.L. has a new recall — straight from official government data. Unsubscribe anytime.