Biomet, Inc. recalls Green Mamba Suture Passer
- Recall date
- December 2, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0472-2016
- FDA classification
- Class II
- Brand / firm
- Biomet, Inc.
- Sold / distributed
- FL, NJ, WI, UT, TX, KY, KS, OH, GA, LA, AL, NY, NC, AR, MO, VA, CA, MN, IN, HI
Why it was recalled
There may be a separation and an insufficient weld between the spring and washer. The weld between the spring and washer may fracture, causing the Mamba Suture Passer instruments jaw to not actuate properly. 6 complaints have been received in which 3 have been reported with a delay up to 10 minutes.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Green Mamba Suture Passer