Medical device recalls Moderate risk

Biomet, Inc. recalls Zimmer Biomet Comprehensive Segmental Revision System (SRS). Orthopedic implants for Proximal Humeral Reconstruction, D…

Recall date
February 28, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1173-2018
FDA classification
Class II
Brand / firm
Biomet, Inc.
Sold / distributed
Distribution US Nationwide

Why it was recalled

During assembly, an SRS Humeral Distal Body implant failed to engage correctly at the point of reduction with a Discovery Elbow System Humeral Condyle. A mismatch between the screw holes in the Humeral Body & Elbow Condyle affects assembly during surgery. If the product has already been implanted, no action is required. The mismatch is estimated to prevent screw engagement in less than 1% of cases.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Zimmer Biomet Comprehensive Segmental Revision System (SRS). Orthopedic implants for Proximal Humeral Reconstruction, Distal Humeral Reconstruction, and Total Humeral Reconstruction

Get recall alerts

Free email alert whenever Biomet, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Biomet, Inc.