Why it was recalled
During assembly, an SRS Humeral Distal Body implant failed to engage correctly at the point of reduction with a Discovery Elbow System Humeral Condyle. A mismatch between the screw holes in the Humeral Body & Elbow Condyle affects assembly during surgery. If the product has already been implanted, no action is required. The mismatch is estimated to prevent screw engagement in less than 1% of cases.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Zimmer Biomet Comprehensive Segmental Revision System (SRS). Orthopedic implants for Proximal Humeral Reconstruction, Distal Humeral Reconstruction, and Total Humeral Reconstruction