Biomet Spine, LLC recalls Telluride Percutaneous Rod Inserter Long, Model #7706-1904. The Inserter is a reusable device provided within an alumin…
- Recall date
- December 19, 2011
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1089-2015
- FDA classification
- Class II
- Brand / firm
- Biomet Spine, LLC
- Sold / distributed
- Distributed to TX, OH, NE and NY.
Why it was recalled
The Telluride Percutaneous Rod Inserter Long does not contain a feature that locks the Rod in place during use. There is a potential for the Rod to disengage from the Inserter during insertion, and could lead to harm to the patient.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Telluride Percutaneous Rod Inserter Long, Model #7706-1904. The Inserter is a reusable device provided within an aluminum sterilization case along with other orthopedic manual surgical instruments. The case consists of three trays of instruments, and the middle tray contains a quantity of two Inserters. The Inserter is for use as part of a Telluride System pedicle screw procedure.