Medical device recalls Moderate risk

Biomet Spine, LLC recalls Telluride Percutaneous Rod Inserter Long, Model #7706-1904. The Inserter is a reusable device provided within an alumin…

Recall date
December 19, 2011
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1089-2015
FDA classification
Class II
Brand / firm
Biomet Spine, LLC
Sold / distributed
Distributed to TX, OH, NE and NY.

Why it was recalled

The Telluride Percutaneous Rod Inserter Long does not contain a feature that locks the Rod in place during use. There is a potential for the Rod to disengage from the Inserter during insertion, and could lead to harm to the patient.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Telluride Percutaneous Rod Inserter Long, Model #7706-1904. The Inserter is a reusable device provided within an aluminum sterilization case along with other orthopedic manual surgical instruments. The case consists of three trays of instruments, and the middle tray contains a quantity of two Inserters. The Inserter is for use as part of a Telluride System pedicle screw procedure.