Boston Scientific Corp recalls IntellaTip MiFi XP Asymmetric (N4) Curve 7/110/2.5/8-10 ; Material Number: M004PM4790N40; Catalog number: PM4790N4; Car…
- Recall date
- November 20, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0465-2016
- FDA classification
- Class II
- Brand / firm
- Boston Scientific Corp
- Sold / distributed
- Worldwide Distribution - US Nationwide in the state of: PA, FL, IN, MI, TN, AL, IL, KY, MI, TX, OH, MO, MA, MN, ID, AL, TN, VA, CA and internationally to Austria, Australia, Belgium, Switzerland, Germany, Great Britain, Greece, Hong Kong, Israel, Iran, Italy, Japan, Kuwait, Latvia, Malaysia, Nether…
Why it was recalled
During routine monitoring of complaints, BSC noted that the IntellaTip MiFi XP N4 complaint rate for distal tip bends/kinks was above the anticipated rate in BSC's product risk documentation.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
IntellaTip MiFi XP Asymmetric (N4) Curve 7/110/2.5/8-10 ; Material Number: M004PM4790N40; Catalog number: PM4790N4; Cardiac: The IntellaTip MiFi XP Catheter is indicated for interruption of accessory atrioventricular conduction pathways associated with tachycardia, for treatment of AV nodal reentrant tachycardia, for treatment of atrial flutter tachycardia, and for creation of complete AV block in patients with a rapid ventricular response to an atrial arrhythmia typically chronic, drug refractory atrial fibrillation.
Get recall alerts
Free email alert whenever Boston Scientific Corp has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Boston Scientific Corp