Boston Scientific Corporation recalls AURIGA 30 GENERAL SYSTEM UPN: M0068S30G0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System…
- Recall date
- April 8, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1397-2019
- FDA classification
- Class II
- Brand / firm
- Boston Scientific Corporation
- Sold / distributed
- Worldwide distribution to Australia, Belgium, Brazil, Chile, Colombia, Egypt, France, Germany, Great Britain, India, Indonesia, Iran, Italy, Lebanon, Malaysia, Mexico, Romania, Russia Fed., South Africa, South Korea, Spain, Taiwan, United Arab Emirates and Vietnam
Why it was recalled
Certain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console is designed to stop emitting laser pulses and enters a stand-by mode after a low power error message.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
AURIGA 30 GENERAL SYSTEM UPN: M0068S30G0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery.