Habib EUS RFA recalled over fire hazard
- Recall date
- November 18, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Boston Scientific Corporation recalls Habib EUS RFA, 10mm electrode active tip, Material Number/UPN: 6500 - Product Usage: The Habib EUS RFA is indicated for…
- Recall number
- Z-0667-2020
- FDA classification
- Class II
- Brand / firm
- Boston Scientific Corporation
- Sold / distributed
- Worldwide distribution - US Nationwide distribution in the states of AL, AZ, CA, CO, CT, FL, NJ, NY, MA, MD, MN, TN, TX and countries of Canada, India, Australia, Great Britain, Greece, Spain.
Why it was recalled
During testing on the Habib EUS RFA device, a temperature increase was observed at the tip of the EUS-FNA needle. It was determined that capacitive 5 coupling presents a risk of energy transfer from the Habib EUS RFA active tip to the introducer device. This capacitive coupling could result in thermal injury to tissue outside of the targeted treatment area due to its use with an un-insulated EUS-FNA needle requiring little or no medical intervention.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Habib EUS RFA, 10mm electrode active tip, Material Number/UPN: 6500 - Product Usage: The Habib EUS RFA is indicated for coagulation and ablation of soft tissue when used in conjunction with compatible radiofrequency generator.