Boston Scientific Corporation recalls Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) System, labeled as EMBLEMTM S-ICD Pulse Generator
- Recall date
- August 21, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2542-2019
- FDA classification
- Class II
- Brand / firm
- Boston Scientific Corporation
- Sold / distributed
- Nationwide distribution to Alabama, Arizona, Connecticut, Delaware, Florida, Georgia, Hawaii, Illinois, Indiana, Iowa, Kansas, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, New Hampshire, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, O…
Why it was recalled
Potential need for device replacement - Elective Replacement Indicator (ERI) or End of Life (EOL) - earlier than expected due to compromised performance of an electrical component causing accelerated battery depletion.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) System, labeled as EMBLEMTM S-ICD Pulse Generator