The LinearTM 3-4 70 cm 8 Contact Lead Kit recalled over sterility concerns
- Recall date
- April 24, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Boston Scientific Neuromodulation Corporation recalls The LinearTM 3-4 70 cm 8 Contact Lead Kit, part number M365SC2352700 is part of a spinal cord stimulation system for pr…
- Recall number
- Z-2122-2020
- FDA classification
- Class II
- Brand / firm
- Boston Scientific Neuromodulation Corporation
- Sold / distributed
- Distributed OUS only. to 14 countries.
Why it was recalled
The firm is initiating a voluntary removal of certain spinal cord stimulation leads, lead extensions, adapters, and deep brain stimulation lead extensions due to the "use by date" displayed on the label which is later than the correct "use by date". No adverse health consequence is expected to occur from this issue.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
The LinearTM 3-4 70 cm 8 Contact Lead Kit, part number M365SC2352700 is part of a spinal cord stimulation system for prescription use only as an aid in the management of chronic intractable pain of the trunk and/or limbs. Package Contents Percutaneous Permanent Lead Kits (1) Percutaneous Lead with pre-loaded Curved Stylet (1) Stylet Ring with a Curved and a Straight Stylet (4) Suture Sleeves (1) Insertion Needle with Stylet (1) Lead Blank (1) Steering Cap (1) OR Cable Assembly (2) Lead Position Labelsleft and right (non-sterile) (1) Device Registration Form/Temporary Patient Identification Card (1) Manual