Bound Tree Medical, LLC recalls curaplex Epi Safe Administration and Training Kits, #8600-01100, Contains: 2 Epi Safe Administration Kit (8600-01101),…
- Recall date
- November 2, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- D-0383-2019
- FDA classification
- Class II
- Brand / firm
- Bound Tree Medical, LLC
- Sold / distributed
- Nationwide in the USA
Why it was recalled
Labeling Incorrect Instructions: This recall has been initiated because the Instructions for Use included in the Epi-Safe Kits for the Epi-Safe Syringe (and/or the Safety Lok Syringe in Model 8600-01120) recommend a midpoint dosage of epinephrine for children between 0 lbs. and 66 lbs. of 0.15mL that has not been approved by FDA for children. Reliance on the current IFU and/or use of the Epi-Safe Syringe as instructed may result in the administration of an inappropriate dosage of epinephrine to young children experiencing anaphylaxis, especially those under 33 lbs. Additionally the Epi Safe syringe does not have the correct markings to allow a health care provider to draw up the correct dosage for children under 33 kg.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
curaplex Epi Safe Administration and Training Kits, #8600-01100, Contains: 2 Epi Safe Administration Kit (8600-01101), 1 Epi Safe Training Kit (8600-01102), Rx Only, Distributed by Sarnova HC, LLC s family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc. 5000 Tuttle Crossing Blvd. Dublin, OH 43016
Get recall alerts
Free email alert whenever Bound Tree Medical, LLC has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Bound Tree Medical, LLC