Drug & medication recalls Moderate risk

Bound Tree Medical, LLC recalls curaplex Epi Safe Kit, 8600-01101, contains:" 1ml Vial of Epinephrine " 1 Epi-Safe Syringe " 1 Safety Needle " 2 Alcoho…

Recall date
November 2, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
D-0384-2019
FDA classification
Class II
Brand / firm
Bound Tree Medical, LLC
Sold / distributed
Nationwide in the USA

Why it was recalled

Labeling Incorrect Instructions: This recall has been initiated because the Instructions for Use included in the Epi-Safe Kits for the Epi-Safe Syringe (and/or the Safety Lok Syringe in Model 8600-01120) recommend a midpoint dosage of epinephrine for children between 0 lbs. and 66 lbs. of 0.15mL that has not been approved by FDA for children. Reliance on the current IFU and/or use of the Epi-Safe Syringe as instructed may result in the administration of an inappropriate dosage of epinephrine to young children experiencing anaphylaxis, especially those under 33 lbs. Additionally the Epi Safe syringe does not have the correct markings to allow a health care provider to draw up the correct dosage for children under 33 kg.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

curaplex Epi Safe Kit, 8600-01101, contains:" 1ml Vial of Epinephrine " 1 Epi-Safe Syringe " 1 Safety Needle " 2 Alcohol Prep Pads " 1 Adhesive Dressing " 1 Insert, Rx Only. Distributed by Sarnova HC, LLC s family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc. 5000 Tuttle Crossing Blvd. Dublin, OH 43016

Get recall alerts

Free email alert whenever Bound Tree Medical, LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Bound Tree Medical, LLC