Brainlab AG recalls Patient Data Manager, Model/Catalogue Numbers: 25100-09 - PATIENT DATA MANAGER 2.1.1LICENSE ONLY 25100-10 - PATIENT DAT…
- Recall date
- September 25, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0176-2020
- FDA classification
- Class II
- Brand / firm
- Brainlab AG
- Sold / distributed
- worldwide
Why it was recalled
The Brainlab Patient Data Manager software (PDM) in combination with the Brainlab iPlan CMF/Cranial/ENT/Spine software 2.1.1, 2.2.0, 2.2.1, 2.3.0, 2.3.1, 2.4.0, 2.5.0, 2.5.1 (if used in combination with iPlan CMF/Cranial/ENT/Spine 3.0.5, 3.0.6) that may lead to shifts applied to segmentation objects in iPlan being not reflected when exported into DICOM under specific circumstances.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Patient Data Manager, Model/Catalogue Numbers: 25100-09 - PATIENT DATA MANAGER 2.1.1LICENSE ONLY 25100-10 - PATIENT DATA MANAGER 2.2.0 LICENSE ONLY 25100-11 - PATIENT DATA MANAGER 2.2.1 LICENSE ONLY 25100-12 - PATIENT DATA MANAGER 2.3 INSTALLER 25100-12A - PATIENT DATA MANAGER 2.3.1 INSTALLER 25100-13 - PATIENT DATA MANAGER 2.3 LICENSE 25100-12B - PATIENT DATA MANAGER 2.4.0 INSTALLER 25100-13A - PATIENT DATA MANAGER 2.4 LICENSE 25100-12C - PATIENT DATA MANAGER 2.5.0 INSTALLER 25100-12D - PATIENT DATA MANAGER 2.5.1 INSTALLER 25100-13B - PATIENT DATA MANAGER 2.5 LICENSE used in combination with iPlan CMF/Cranial/ENT/Spine 3.0.5, 3.0.6. Product Usage: Patient Data Manager is intended to load patient data from and export patient data to connected network locations (like file shares and PACS) and removable media. Furthermore, the software is intended to start other applications (like Brainlab planning and navigation applications) and to route applications and connected video sources to displays attached to the system that are running Patient Data Manager. The software is also intended to connect two Brainlab systems to each other via network.
Get recall alerts
Free email alert whenever Brainlab AG has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Brainlab AG