Medical device recalls Moderate risk

C.R. Bard, Inc. recalls BARDEX I.C. Anti-Infective 2-Way 3cc Foley Catheter, Rx only. Intended for urinary bladder drainage in female patients…

Recall date
December 18, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1357-2016
FDA classification
Class II
Brand / firm
C.R. Bard, Inc.
Sold / distributed
Distributed US (nationwide) and the countries of Canada and Australia.

Why it was recalled

Labeling error: Catalog contains Instructions for Use (IFU) error; incorrect inflation volume.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

BARDEX I.C. Anti-Infective 2-Way 3cc Foley Catheter, Rx only. Intended for urinary bladder drainage in female patients requiring catheterization for management of incontinence, voiding dysfunction and surgical procedure.