Canon Medical System, USA, INC. recalls Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is use…
- Recall date
- February 8, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0360-2020
- FDA classification
- Class II
- Brand / firm
- Canon Medical System, USA, INC.
- Sold / distributed
- Worldwide distribution - US Nationwide distribution in the states of AL, CT, FL, IL, OR, PA, and TX; and country of Netherlands.
Why it was recalled
Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.
Get recall alerts
Free email alert whenever Canon Medical System, USA, INC. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Canon Medical System, USA, INC.