Medical device recalls Moderate risk

Cardinal Health 200, LLC recalls Genius 2 Tympanic Thermometer OEM Tympanic Item Code:303062

Recall date
October 28, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0992-2020
FDA classification
Class II
Brand / firm
Cardinal Health 200, LLC
Sold / distributed
Worldwide Distribution - US Nationwide and Foreign in the countries of: Australia Austria Belgium Canada Canary Islands Chile Costa Rica Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France French Guiana French Polynesia Gabon Germany Gibraltar Greece Guadeloupe Hong Kong Iran, Islami…

Why it was recalled

The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2¿C for Genius 2 and ¿ 0.3¿C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2¿C for Genius 2 or +0.3¿C for Genius 3.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Genius 2 Tympanic Thermometer OEM Tympanic Item Code:303062