Cardinal Health 200, LLC recalls Genius 2 Tympanic Thermometer OEM Tympanic Item Code: 303063
- Recall date
- October 28, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0993-2020
- FDA classification
- Class II
- Brand / firm
- Cardinal Health 200, LLC
- Sold / distributed
- Worldwide Distribution - US Nationwide and Foreign in the countries of: Australia Austria Belgium Canada Canary Islands Chile Costa Rica Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France French Guiana French Polynesia Gabon Germany Gibraltar Greece Guadeloupe Hong Kong Iran, Islami…
Why it was recalled
The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2¿C for Genius 2 and ¿ 0.3¿C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2¿C for Genius 2 or +0.3¿C for Genius 3.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Genius 2 Tympanic Thermometer OEM Tympanic Item Code: 303063
Get recall alerts
Free email alert whenever Cardinal Health 200, LLC has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Cardinal Health 200, LLC