Cocoon Convective Warming System recalled over fire hazard
- Recall date
- January 7, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Care Essentials Pty., Ltd. recalls Cocoon Convective Warming System, Product Code: CWS4000 (110V).
- Recall number
- Z-0986-2020
- FDA classification
- Class II
- Brand / firm
- Care Essentials Pty., Ltd.
- Sold / distributed
- US Nationwide in the states of AZ CA FL GA KS LA MA MD MI MN NC NM NY OH PA QC SC TX VA WV and the Foreign distribution to Canada.
Why it was recalled
Potential for thermal damage near the power cord connector and Power Entry Module (PEM) due to improper connection of the power cord connector and PEM pins on the Cocoon Convective Warming System.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Cocoon Convective Warming System, Product Code: CWS4000 (110V).