Medical device recalls Moderate risk

Carefusion 2200 Inc recalls (1) Carefusion V. Mueller Graves Vaginal Speculum, Catalog #GL11, Medium Blade, 4 x 1 3/8" (10.2 X 3.5cm); and (2) Care…

Recall date
March 16, 2023
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1421-2023
FDA classification
Class II
Brand / firm
Carefusion 2200 Inc
Sold / distributed
Distribution was made to AL, AR, AZ, CA, CO, CT, FL, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, VA, WA, WI, and WV, including Puerto Rico. There was also government/military distribution. Foreign distribution was made to Canada.

Why it was recalled

A portion of the two products were incorrectly marked (etched) and packaged resulting in a product mismatch. The labels and laser etchings were switched on the two products.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

(1) Carefusion V. Mueller Graves Vaginal Speculum, Catalog #GL11, Medium Blade, 4 x 1 3/8" (10.2 X 3.5cm); and (2) Carefusion V. Mueller Pederson Vaginal Specula, Catalog #GL31, Medium, 4 x 7/8" (10.2 x 2.2cm).