Medical device recalls Moderate risk

CareFusion 303, Inc. recalls Alaris System PC unit, model no. 8000 and 8015. The central programming, monitoring and power supply component for the…

Recall date
November 1, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1520-2017
FDA classification
Class II
Brand / firm
CareFusion 303, Inc.
Sold / distributed
Worldwide distribution. US nationwide, Europe, Australia, New Zealand, South Africa, Greater Asia, Middle East, and Canada.

Why it was recalled

Reports where the Low Battery alarm and/or the Very Low Battery alarm are not being triggered before the battery is discharged and all infusion channels are stopped.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Alaris System PC unit, model no. 8000 and 8015. The central programming, monitoring and power supply component for the Alaris infusion pump System.