CAREFUSION recalls BD Gravity Burette Set 60 DROP 1.2 MF 2 SmartSite VALVES BALL VALVE DRIP CHAMBER V/NV, REF: 10037031; BD Gravity Burett…
- Recall date
- August 2, 2021
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2456-2021
- FDA classification
- Class II
- Brand / firm
- CAREFUSION
- Sold / distributed
- Worldwide distribution - US Nationwide distribution in the states of FL, PA, NC, OH, SC, IL, RI, MO, NY, TN, AL, VA, PR, CA, NE, NJ, MI, MD, NM, LA, KY, NV, ME, OK, WA, MA, WV, TX, AR, OR, KS, GA, VT, IN, WY, AZ, CO, MN, WI, UT, SD, MS, ID, MT, ND, HI, IA, DC, AK, NH, CT, DE, GU and the countries o…
Why it was recalled
Needle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
BD Gravity Burette Set 60 DROP 1.2 MF 2 SmartSite VALVES BALL VALVE DRIP CHAMBER V/NV, REF: 10037031; BD Gravity Burette Set 20 DROP 1.2 Micron Filter 2 SmartSite VALVES BALL VALVE DRIP CHAMBER V/NV , REF: 10037032; BD GRAVITY BURETTE SET 60 DROP 2 Smartsite VALVES BALL DRIP CHAMBER \VENTED/NONVENTED, REF: 82104E; BD ADD-ON BURETTE SET Smartsite VALVE, REF: 82113E; BD ADD-ON BURETTE SET Smartsite VALVE VENTED/NONVENTED, REF: 82114E; BD ADD-ON BURETTE SET Smartsite VALVE BALL VALVE DRIP CHAMBER VENTED/NONVENTED, REF: 82115E