Medical device recalls Moderate risk

CAREFUSION recalls BD Gravity Burette Set 60 DROP 1.2 MF 2 SmartSite VALVES BALL VALVE DRIP CHAMBER V/NV, REF: 10037031; BD Gravity Burett…

Recall date
August 2, 2021
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2456-2021
FDA classification
Class II
Brand / firm
CAREFUSION
Sold / distributed
Worldwide distribution - US Nationwide distribution in the states of FL, PA, NC, OH, SC, IL, RI, MO, NY, TN, AL, VA, PR, CA, NE, NJ, MI, MD, NM, LA, KY, NV, ME, OK, WA, MA, WV, TX, AR, OR, KS, GA, VT, IN, WY, AZ, CO, MN, WI, UT, SD, MS, ID, MT, ND, HI, IA, DC, AK, NH, CT, DE, GU and the countries o…

Why it was recalled

Needle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

BD Gravity Burette Set 60 DROP 1.2 MF 2 SmartSite VALVES BALL VALVE DRIP CHAMBER V/NV, REF: 10037031; BD Gravity Burette Set 20 DROP 1.2 Micron Filter 2 SmartSite VALVES BALL VALVE DRIP CHAMBER V/NV , REF: 10037032; BD GRAVITY BURETTE SET 60 DROP 2 Smartsite VALVES BALL DRIP CHAMBER \VENTED/NONVENTED, REF: 82104E; BD ADD-ON BURETTE SET Smartsite VALVE, REF: 82113E; BD ADD-ON BURETTE SET Smartsite VALVE VENTED/NONVENTED, REF: 82114E; BD ADD-ON BURETTE SET Smartsite VALVE BALL VALVE DRIP CHAMBER VENTED/NONVENTED, REF: 82115E

Get recall alerts

Free email alert whenever CAREFUSION has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: CAREFUSION