Medical device recalls Moderate risk

CAREFUSION recalls BD Syringe Set 20 DP MIC TBG 0.2MF CKV Anti-Siphon SmartSite VLV Sensing Disc, REF: 10798696; BD Syringe Set 20 DP MIC…

Recall date
August 2, 2021
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2457-2021
FDA classification
Class II
Brand / firm
CAREFUSION
Sold / distributed
Worldwide distribution - US Nationwide distribution in the states of FL, PA, NC, OH, SC, IL, RI, MO, NY, TN, AL, VA, PR, CA, NE, NJ, MI, MD, NM, LA, KY, NV, ME, OK, WA, MA, WV, TX, AR, OR, KS, GA, VT, IN, WY, AZ, CO, MN, WI, UT, SD, MS, ID, MT, ND, HI, IA, DC, AK, NH, CT, DE, GU and the countries o…

Why it was recalled

Needle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

BD Syringe Set 20 DP MIC TBG 0.2MF CKV Anti-Siphon SmartSite VLV Sensing Disc, REF: 10798696; BD Syringe Set 20 DP MIC TBG CKV Anti-Siphon SmartSite VLV Sensing Disc, REF: 10798703;