Carl Zeiss Meditec AG recalls Carl Zeiss Meditec AG, Goeschwitzer Strasse 51-52 07745 Jena, Germany, IOLMaster 700, REF 1932-169 SN 1185393 2017-09-01
- Recall date
- June 25, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0108-2019
- FDA classification
- Class II
- Brand / firm
- Carl Zeiss Meditec AG
- Sold / distributed
- US: CA (Mountain View and San Jose) A single system. No distributed OUS.
Why it was recalled
Calibration setting was incorrectly entered during manufacturing causing the optical coherence tomography (OCT) to obtain the incorrect two-dimensional images of ocular structures of the anterior and posterior segments of the eye.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Carl Zeiss Meditec AG, Goeschwitzer Strasse 51-52 07745 Jena, Germany, IOLMaster 700, REF 1932-169 SN 1185393 2017-09-01
Get recall alerts
Free email alert whenever Carl Zeiss Meditec AG has a new recall — straight from official government data. Unsubscribe anytime.