Medical device recalls Moderate risk

Carl Zeiss Meditec AG recalls Carl Zeiss Meditec Ag VISULENS 550 REF 2268-038 - Product Usage: is a device for measuring the refractive power of cont…

Recall date
August 6, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0613-2021
FDA classification
Class II
Brand / firm
Carl Zeiss Meditec AG
Sold / distributed
Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, MI, MN, NC, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TX, UT, VA, WA and WY. The countries of Argentina, Austria, Australia, Belgium, Bolivia,…

Why it was recalled

Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Carl Zeiss Meditec Ag VISULENS 550 REF 2268-038 - Product Usage: is a device for measuring the refractive power of contact lenses and spectacle lenses and displaying measurement values for sphere, cylinder and prism.

Get recall alerts

Free email alert whenever Carl Zeiss Meditec AG has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Carl Zeiss Meditec AG