Medical device recalls Moderate risk

Carl Zeiss Meditec, Inc. recalls VisuMax Software Version 2.10.13 with activated Software-Module ReLEx (SMILE) option used in conjunction with the VisuM…

Recall date
December 6, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0680-2020
FDA classification
Class II
Brand / firm
Carl Zeiss Meditec, Inc.
Sold / distributed
US nationwide distribution.

Why it was recalled

VisuMax devices with software version 2.10.13 and activated SW-Module ReLEx (SMILE) option may experience a software issue. In case of a suction loss during treatment the software allows the user to choose the option immediate restart or restart treatment. Suction loss can occur in phase 1 through phase 5. The software defect refers to Phase 2 (between 10% and 100% of lower lenticule cut) only. The software offers a flap cut, but due to the software issue it performs a cap cut instead of a flap cut if the user proceeds.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

VisuMax Software Version 2.10.13 with activated Software-Module ReLEx (SMILE) option used in conjunction with the VisuMax Femtosecond Laser System for Refractive Correction. Catalog Number 0000000-1345-518 Product Usage: The VisuMax Femtosecond Laser is a precision ophthalmic surgical laser designed for the creation of incisions in the cornea. The VisuMax Femtosecond Laser is indicated for use in small incision lenticule extraction (SMILE) for the reduction or elimination of myopia. This device is intended for use in small incision lenticule extraction (SMILE) for the reduction or elimination of myopia >=-1.00 D to <=-8.00 D, with <=-0.50 D cylinder and MRSE <=-8.25 D in the eye to be treated in patients who are 22 years of age or older with documentation of stable manifest refraction over the past year as demonstrated by a change of <=0.50 D MRSE.

Get recall alerts

Free email alert whenever Carl Zeiss Meditec, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Carl Zeiss Meditec, Inc.