Carl Zeiss Metrology Inc recalls Metrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components made of plastic…
- Recall date
- November 14, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0379-2018
- FDA classification
- Class II
- Brand / firm
- Carl Zeiss Metrology Inc
- Sold / distributed
- Worldwide Distribution - US, including the states of : WI, NC, GA, IL, AZ, FL, CA, NJ, and territory of PR; and the country of Canada.
Why it was recalled
Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the control panel.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Metrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components made of plastic or light metal.
Get recall alerts
Free email alert whenever Carl Zeiss Metrology Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Carl Zeiss Metrology Inc