Medical device recalls Moderate risk

Centurion Medical Products Corporation recalls Centurion CIRCLAMP W/1.1CM BELL - single-use, disposable instrument used to compress the foreskin of the penis during c…

Recall date
August 28, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0120-2020
FDA classification
Class II
Brand / firm
Centurion Medical Products Corporation
Sold / distributed
Nationwide

Why it was recalled

A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Centurion CIRCLAMP W/1.1CM BELL - single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant. ReOrder: 310CR

Get recall alerts

Free email alert whenever Centurion Medical Products Corporation has a new recall — straight from official government data. Unsubscribe anytime.