Why it was recalled

Cepheid has received reports that some of our customers have experienced a higher number of temperature drift errors than would typically be expected when using this test and its intended use specimen types. Investigation of the issue also revealed an occasional probability of false negative results which remains consistent with package insert claims.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Xpert CT/NG Catalog GXCT/NG-120 Batch 1000049568, Catalog GXCT/NG-10 Batch 1000053075, Catalog GXCT/NG-10 Batch 1000056156 and Catalog GXCT/NG-10 Batch 1000056157. Is a qualitative in vitro real-time PCR test.