Clinical Diagnostic Solutions, Inc. recalls Medonic M-Series Hematology Analyzer M16S BD ABR US Product code: 1400075; OUS Product code 1400011
- Recall date
- January 2, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1001-2020
- FDA classification
- Class II
- Brand / firm
- Clinical Diagnostic Solutions, Inc.
- Sold / distributed
- Worldwide distribution. US Nationwide, Venezuela, Africa, Italy, Brazil, and Ukraine. No governmental, Canadian, or Mexican consignees.
Why it was recalled
A mix-up of autosampler tube positions with a possibility of misidentification of sample results and a risk of operator injury from an exposed aspiration needle.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Medonic M-Series Hematology Analyzer M16S BD ABR US Product code: 1400075; OUS Product code 1400011