Clinical Innovations, LLC recalls Clearview Total Uterine Manipulator (model UM750), The device is labeled in part, "ClearView(R) TOTAL, 7 cm* * *CLINICA…
- Recall date
- May 30, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2118-2020
- FDA classification
- Class II
- Brand / firm
- Clinical Innovations, LLC
- Sold / distributed
- International distribution to the country Japan.
Why it was recalled
An incorrect component used during manufacturing. The uterine manipulator tip support does not have the "notch" required to attach the spacer and or colpotomy cup. This may result in a delay in surgery and additional surgical interventions to completed the procedure. There was a report of an injury.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Clearview Total Uterine Manipulator (model UM750), The device is labeled in part, "ClearView(R) TOTAL, 7 cm* * *CLINICAL INNOVATIONS Manufacturer: Clinical Innovations, LLC* * *Reorder Number: UM750* * *LOT 181183* * *." Each device is accompanied by Instruction for Use which is similarly labeled.