Medical device recalls Moderate risk

Clinical Innovations, LLC recalls Clearview Total Uterine Manipulator (model UM750), The device is labeled in part, "ClearView(R) TOTAL, 7 cm* * *CLINICA…

Recall date
May 30, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2118-2020
FDA classification
Class II
Brand / firm
Clinical Innovations, LLC
Sold / distributed
International distribution to the country Japan.

Why it was recalled

An incorrect component used during manufacturing. The uterine manipulator tip support does not have the "notch" required to attach the spacer and or colpotomy cup. This may result in a delay in surgery and additional surgical interventions to completed the procedure. There was a report of an injury.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Clearview Total Uterine Manipulator (model UM750), The device is labeled in part, "ClearView(R) TOTAL, 7 cm* * *CLINICAL INNOVATIONS Manufacturer: Clinical Innovations, LLC* * *Reorder Number: UM750* * *LOT 181183* * *." Each device is accompanied by Instruction for Use which is similarly labeled.

Get recall alerts

Free email alert whenever Clinical Innovations, LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Clinical Innovations, LLC