Medical device recalls Moderate risk

Collagen Matrix Inc recalls NuOss XC Sinus Reference No 509-3001 (BMCU17, size 17 mm diameter x 10 mm height) Reference No 509-3002 (BMCU22, size 2…

Recall date
May 13, 2013
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1097-2016
FDA classification
Class II
Brand / firm
Collagen Matrix Inc
Sold / distributed
US Distribution to state of: MA.

Why it was recalled

During the 36 month real time shelf life testing of NuOss XC Sinus (BMCU-Umbrella) to verify the 3 year shelf life designated for the device, 3 of 13 products did not pass the recovery test, which requires that the compressed product fully expand within 1 minute of hydration time.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

NuOss XC Sinus Reference No 509-3001 (BMCU17, size 17 mm diameter x 10 mm height) Reference No 509-3002 (BMCU22, size 22 mm diameter x 12 m height) Manufacturer:- Collagen Matrix, Distributor:- ACE Surgical Supply Co, Inc, Intended use for dental surgery such as -augmentation or reconstructive treatment of alveoloar ridge -filling of infrabony periodontal defects -filling of defects after root resection, apicoectomy and cystectomy -filling of extraction sockets to enhance preservation of the alveolar ridge -elevation of maxillary sinus floor -filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR) -Filling of pre-implant defects in conjunction with products intended for Guided Bone Regeneration

Get recall alerts

Free email alert whenever Collagen Matrix Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Collagen Matrix Inc