Medical device recalls Moderate risk

Conformis, Inc. recalls Conformis iTotal Hip Replacement System: HBS-033-00xx-010101, (HIP-PATIENT SPECIFIC STEM LEFT)

Recall date
March 29, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1775-2019
FDA classification
Class II
Brand / firm
Conformis, Inc.
Sold / distributed
Nationwide

Why it was recalled

Discrepancy of 3mm between the values displayed and approved by surgeons in iViews as the neck/head lengths and the resulting neck/head lengths in the manufactured implants.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Conformis iTotal Hip Replacement System: HBS-033-00xx-010101, (HIP-PATIENT SPECIFIC STEM LEFT)

Get recall alerts

Free email alert whenever Conformis, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Conformis, Inc.