Medical device recalls Moderate risk

Conmed Corporation recalls Pediatric ThermoGard¿ Dual Dispersive Electrodes (for patients 5-15 kg.), 10' (3.05m) Cable, Catalog Number 51-7710. In…

Recall date
May 18, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2597-2016
FDA classification
Class II
Brand / firm
Conmed Corporation
Sold / distributed
Worldwide Distribution -- USA, including AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, & Puerto Rico; and, the countries of Australia, Canada, Belgium, Brunei Darussalam…

Why it was recalled

A version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Pediatric ThermoGard¿ Dual Dispersive Electrodes (for patients 5-15 kg.), 10' (3.05m) Cable, Catalog Number 51-7710. Intended to be used for the dispersion and return to the electrosurgical generator.

Get recall alerts

Free email alert whenever Conmed Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Conmed Corporation