Sterile Handpiece Sheaths recalled over sterility concerns
- Recall date
- February 28, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- ConMed Corporation recalls Sterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000 Electrosur…
- Recall number
- Z-1759-2017
- FDA classification
- Class II
- Brand / firm
- ConMed Corporation
- Sold / distributed
- Worldwide Distribution - US (nationwide) Internationally to Australia, Canada, Germany, Italy, UK, Great Britain, and Ireland.
Why it was recalled
For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were sterile.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Sterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000 Electrosurgical Unit (ESU) cartons, 230 VOLTS AC, REF/Catalog Number 7-900-230, Rx Only. There is no problem with the Hyfrecator 2000 Electrosurgical Units. -- ASSEMBLED IN MEXICO -- CONMED CORPORATION 525 FRENCH ROAD UTICA, NY 13502