Medical device recalls Moderate risk

Contour Fabricators Inc recalls Zero-Gravity Radiation Protection System Radiology Personnel protective shield.

Recall date
September 9, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0080-2017
FDA classification
Class II
Brand / firm
Contour Fabricators Inc
Sold / distributed
Nationwide. Canada. AUSTRALIA AUSTRIA BELGIUM BRUNEI CHINA FRANCE GERMANY INDA INDONESIA ISRAEL ITALY KSA MALAYSIA NETHERLANDS POLAND QATAR QUEBEC SAUDI ARABIA SINGAPORE SOUTH AFRICA TAIWAN (R.O.C.)

Why it was recalled

Two customer complaints suggest that excessive rotation of the cable in one direction (clockwise or counter-clockwise) without allowing the cable to unwind to its normal condition may cause the cable to break suddenly and the body shield to fall. While no injuries or deaths occurred in either failure, the Zero-GravityTM body shield weighs 55 pounds, and in the event of a cable failure, the body shield will fall immediately. A fall of this nature places the user, associated staff and the patient at risk for a potential injury.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Zero-Gravity Radiation Protection System Radiology Personnel protective shield.

Get recall alerts

Free email alert whenever Contour Fabricators Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Contour Fabricators Inc